Guidelines for Transcutaneous and Pco Measurement
1989

Guidelines for Transcutaneous Measurement of Blood Gases

publication Evidence: moderate

Author Information

Author(s): P. D. Wimberley, R. W. Burnett, A. K. Covington, A. H. J. Maas, O. Mueller-Plathe, O. Siggaard-Andersen, H. F. Weisberg, W. G. Zijstra

Primary Institution: International Federation of Clinical Chemistry (IFCC)

Conclusion

The document provides guidelines for the terminology, methodology, and interpretation of data from transcutaneous Po2 and Pco2 measurements.

Supporting Evidence

  • Transcutaneous measurements are particularly useful in newborns due to their unique physiological characteristics.
  • The guidelines recommend avoiding electrode temperatures above 44 C to prevent skin burns.
  • Calibration of electrodes is crucial for accurate measurements and should be performed regularly.

Takeaway

This document helps doctors understand how to measure oxygen and carbon dioxide levels in the blood using special sensors on the skin, especially for newborns.

Methodology

The guidelines cover the use of heated electrodes for measuring oxygen and carbon dioxide levels in the skin, including calibration and measurement procedures.

Potential Biases

Potential biases include variations in skin anatomy, blood flow, and interference from anesthetic gases.

Limitations

The document does not provide specific data on the accuracy of the measurements or the potential for bias in different patient populations.

Participant Demographics

The guidelines particularly emphasize the application in newborn infants due to their thinner skin and higher capillary density.