A new method for experimental characterisation of scattered radiation in 64-slice CT scanner
2010

New Method for Measuring Scattered Radiation in 64-Slice CT Scanners

publication Evidence: moderate

Author Information

Author(s): Akbarzadeh A, Ay MR, Ghadiri H, Sarkar S, Zaidi H

Primary Institution: Department of Biomedical Imaging, Faculty of Medicine, University of Malaya, Malaysia

Hypothesis

Can a new method using dedicated blockers improve the measurement of scatter radiation in 64-slice CT scanners?

Conclusion

The method developed within this study is applicable to any measurement and is direct with minimal complexity.

Supporting Evidence

  • The method allows for the measurement of scatter radiation in multiple detector rows, improving upon previous single measurement methods.
  • Results showed that increasing tube voltage reduced the scatter-to-primary ratio significantly.
  • The study utilized a 64-slice CT scanner manufactured by GE Healthcare Technologies.

Takeaway

This study found a new way to measure scattered radiation in CT scans, which helps make the images clearer and more accurate.

Methodology

A lead blocker array was installed beneath the collimator of a 64-slice CT scanner to measure scatter profiles and scatter-to-primary ratios.

Limitations

The method may have limitations related to the specific setup and conditions under which measurements were taken.