A comparison between a new automatic system and Doppler method for obtaining ankle brachial pressures
2010

Comparing New and Traditional Methods for Measuring Ankle Brachial Pressures

Sample size: 200 publication Evidence: moderate

Author Information

Author(s): Lewis Jane, Hawkins Melinda, Barree Phillip, Cawley Scott, Dayananda Susan

Primary Institution: Cardiff and Vale University Health Board

Hypothesis

The new device (Dopplex Ability) can perform the ABPI test quicker than Doppler, does not require resting the patient, and provides results in good agreement with Doppler.

Conclusion

The Dopplex Ability provides good agreement with the hand-held Doppler and takes significantly less time, improving the patient experience.

Supporting Evidence

  • The Dopplex Ability showed good agreement with Doppler for both unrested and rested tests.
  • The mean time for Dopplex Ability was significantly less than for Doppler.

Takeaway

This study shows that a new device can check for blood flow problems faster and without needing patients to rest, making it easier for doctors to help people.

Methodology

A randomized cross-over study design was used to compare the new device and Doppler method.

Participant Demographics

200 subjects, 400 limbs, with a focus on those at risk for peripheral arterial disease.

Statistical Information

P-Value

p<0.05

Confidence Interval

95% limits of agreement were +0.22 for unrested and +0.20 for rested Dopplex Ability compared to Doppler.

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1186/1757-1146-3-S1-O15