Anti-CCP: History and its usefulness
2005

Evaluating a New Automated Anti-CCP Test for Rheumatoid Arthritis

Sample size: 543 publication Evidence: moderate

Author Information

Author(s): M. Herold, V. Boeser, E. Russe, W. Klotz

Primary Institution: Innsbruck Medical University

Hypothesis

The new automated method for determining anti-CCP2 will provide comparable results to existing ELISA methods.

Conclusion

The new automated assay for anti-CCP2 demonstrated a diagnostic sensitivity of 80% and specificity of 97%, comparable to existing methods.

Supporting Evidence

  • The new automated assay showed a diagnostic sensitivity of 80% and specificity of 97%.
  • Results from the automated method were comparable to those from the existing ELISA tests.
  • Anti-CCP testing has increasing acceptance in laboratory medicine.

Takeaway

Doctors tested a new machine that checks for a specific antibody in blood to help diagnose rheumatoid arthritis, and it worked just as well as the old tests.

Methodology

Five hundred and forty-three serum samples were tested for anti-CCP2 using both a commercially available ELISA and a new automated method.

Limitations

The study did not address the long-term reliability of the automated method or its performance in diverse populations.

Participant Demographics

395 female and 148 male outpatients from a hospital with joint problems.

Digital Object Identifier (DOI)

10.1080/14767050500134217

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication