Measuring Vitamin D Levels from Blood Spot Cards
Author Information
Author(s): Larkin Emma K., Gebretsadik Tebeb, Koestner Nathan, Newman Mark S., Liu Zhouwen, Carroll Kecia N., Minton Patricia, Woodward Kim, Hartert Tina V.
Primary Institution: Vanderbilt University
Hypothesis
Can blood spot samples accurately measure 25-hydroxyvitamin D levels compared to serum and whole blood samples?
Conclusion
Blood spot samples provide a reliable and less invasive method for measuring vitamin D levels, although they tend to yield slightly lower results than serum or whole blood samples.
Supporting Evidence
- Blood spot measurements were on average 1.1 ng/mL lower than serum measurements.
- 25OHD levels were categorized as sufficient, insufficient, or deficient based on established thresholds.
- Statistical analysis showed a marginal association between dietary vitamin D intake and measured 25OHD levels.
Takeaway
This study shows that we can use small blood samples to check vitamin D levels, which is easier than taking blood from a vein.
Methodology
The study measured 25OHD levels using liquid chromatography–tandem mass spectrometry on samples from blood spot cards, serum, and whole blood from mothers of infants.
Potential Biases
Potential bias due to the lack of detection of 25OHD2 levels and the method of blood spot preparation.
Limitations
The blood spot cards were prepared in the lab and not tested in the field, which may affect results.
Participant Demographics
The study involved 109 biological mothers of infants, with a mean age of 26.5 years, approximately 80% European-American, and 30% smokers.
Statistical Information
P-Value
0.003
Confidence Interval
95% CI: 1.6, 4.8
Statistical Significance
p<0.05