Evaluation of the Jokoo-ION 150 AC Analyser
Author Information
Author(s): Joan Farr6, Carmen Biosca, Romin Galimany
Primary Institution: Servei d'Anitlisis Cliniques, Hospital de Badalona 'Germans Trias Pujol'
Hypothesis
The study aims to evaluate the performance of the Jokoo-ION 150 AC analyser for measuring sodium, potassium, and chloride ion concentrations.
Conclusion
The Jokoo-ION 150 AC analyser demonstrated high precision and acceptable accuracy for measuring sodium, potassium, and chloride ions.
Supporting Evidence
- The within-run imprecision for the Jokoo-ION 150 AC was smaller than 1%.
- The detection limit for sodium is 12.48 mmol/L, for potassium is 0.19 mmol/L, and for chloride is 14.76 mmol/L.
- Carry-over and drift are negligible, smaller than 2%.
- Significant variations due to pH were observed on both sodium and potassium ions.
Takeaway
The Jokoo-ION 150 AC is a machine that helps doctors measure important salts in blood and urine, and it works really well.
Methodology
The study involved evaluating the precision, accuracy, and interference of the Jokoo-ION 150 AC analyser using various control samples and methods.
Potential Biases
Potential biases may arise from the limited experience of laboratories with electrometric techniques.
Limitations
The study did not evaluate the performance of the analyser in all possible clinical scenarios.