The concurrent validity of a portable ultrasound probe for muscle thickness measurements
2025

Validity of a Portable Ultrasound Probe for Measuring Muscle Thickness

Sample size: 18 publication 10 minutes Evidence: moderate

Author Information

Author(s): Homer Kai A., Cross Matt R., Jukic Ivan

Primary Institution: Auckland University of Technology

Hypothesis

Can a portable ultrasound probe accurately measure muscle thickness compared to a stationary device?

Conclusion

The Lumify probe can be used interchangeably with the Vivid S5 for muscle thickness measurements.

Supporting Evidence

  • Systematic bias was low at all sites (≤0.11 cm).
  • Linear associations were found between the devices at each site of measurement (r≥0.95).
  • The difference in muscle thickness between devices was significant only at the anterior thigh (p<0.05).
  • Effect sizes for all sites were considered trivial (ES≤0.131).
  • Proportional bias was detected only at the posterior lower leg.

Takeaway

Researchers tested a new portable ultrasound device to see if it works as well as a bigger, stationary one for measuring muscle thickness, and it does!

Methodology

Participants had their muscle thickness measured at five sites using both the Lumify and Vivid S5 devices in a repeated-measures design.

Potential Biases

Proportional bias was detected at the posterior lower leg, but the absolute differences were trivial.

Limitations

The study only included young, healthy adults engaged in regular resistance training, limiting generalizability.

Participant Demographics

18 participants (9 males, 9 females) aged 25.44 years on average.

Statistical Information

P-Value

p<0.05

Confidence Interval

95% CI provided for various measurements.

Statistical Significance

p<0.05