Measurement of Amniotic Fluid Interleukin-6 Using Commercial Kits
1997

Measuring Interleukin-6 in Amniotic Fluid

Sample size: 8 publication Evidence: moderate

Author Information

Author(s): Tsunenobu Tamura, William W. Andrews, Kelley E. Johnston, G. Philamon Hemstreet, Robert L. Goldenberg

Primary Institution: University of Alabama at Birmingham

Hypothesis

Can commercial kits accurately measure interleukin-6 concentrations in amniotic fluid?

Conclusion

Standardization of the method for measuring IL-6 in amniotic fluid is necessary for consistent laboratory comparisons.

Supporting Evidence

  • IL-6 concentrations in amniotic fluid varied significantly among the kits used.
  • All kits could distinguish higher IL-6 levels in samples from women with chorioamnionitis.
  • Recovery of known IL-6 added to amniotic fluid ranged from 12 to 123%.

Takeaway

This study looked at how well different kits can measure a protein called IL-6 in amniotic fluid, finding big differences between them.

Methodology

Seven commercial kits were used to measure IL-6 in amniotic fluid samples from women with chorioamnionitis and normal pregnancies.

Potential Biases

Differences in antibodies and preparation of standards among kits may introduce bias.

Limitations

The study did not evaluate day-to-day variations for each kit.

Participant Demographics

Samples were taken from three women with clinical chorioamnionitis and five from normal pregnancies.

Statistical Information

Statistical Significance

p<0.05