Tidal Volume Single Breath Washout of Two Tracer Gases - A Practical and Promising Lung Function Test
2011

New Lung Function Test Using Two Tracer Gases

Sample size: 13 publication 10 minutes Evidence: high

Author Information

Author(s): Florian Singer, Georgette Stern, Cindy Thamrin, Oliver Fuchs, Thomas Riedel, Per Gustafsson, Urs Frey, Philipp Latzin

Primary Institution: University Children's Hospital of Bern

Hypothesis

Can a new tidal single breath washout test using sulfur hexafluoride and helium provide accurate measurements of lung function?

Conclusion

The ultrasonic flowmeter accurately measures changes in sulfur hexafluoride and helium washout, indicating that the new test is reliable and repeatable in healthy adults.

Supporting Evidence

  • The USFM reliably measured molar mass during all tests.
  • High association between USFM and mass spectrometer signals was observed.
  • Repeatability and reproducibility of the washout test were confirmed.

Takeaway

Researchers created a new lung test that uses two gases to see how well your lungs work, and it’s quick and easy to do.

Methodology

The study involved 13 healthy adults performing a tidal single breath washout test using a mixture of sulfur hexafluoride and helium, measured with an ultrasonic flowmeter.

Limitations

The study did not include children or patients with lung diseases, limiting the generalizability of the findings.

Participant Demographics

Mean age of participants was 35.2 years, with 7 males.

Statistical Information

P-Value

<0.001

Confidence Interval

−0.0021 to 0.0013 g/mol

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1371/journal.pone.0017588

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication