Detection, measurement, and diagnosis of lung nodules by ultra-low-dose CT in lung cancer screening: a systematic review
2024

Using Ultra-Low-Dose CT to Detect Lung Nodules

Sample size: 1889 publication Evidence: moderate

Author Information

Author(s): Pan Zhijie, Zhang Yaping, Zhang Lu, Wang Lingyun, Zhao Keke, Li Qingyao, Wang Ai, Hu Yanfei, Xie Xueqian

Primary Institution: Shanghai General Hospital, Shanghai Jiao Tong University School of Medicine

Hypothesis

This review aims to summarize the latest advances of ultra-low-dose CT (ULDCT) in the detection, measurement, and diagnosis of lung nodules.

Conclusion

ULDCT proves effective in detecting lung nodules while substantially reducing radiation exposure.

Supporting Evidence

  • Detection rates for lung nodules ranged from 86.1% to 100%.
  • The mean radiation dose for ULDCT was 0.22 ± 0.10 mSv.
  • The diagnosis rate of malignant nodules ranged from 75% to 91%.
  • QUADAS-II showed a generally low risk of bias across studies.
  • Variability of diameter measurements ranged from 2.1% to 14.4%.

Takeaway

Ultra-low-dose CT can find lung nodules without giving patients too much radiation, making it safer for screening.

Methodology

A systematic review of studies in PubMed and Web of Science was conducted, focusing on ULDCT and lung nodules published in the last 5 years.

Potential Biases

The risk of bias was assessed using the QUADAS-II tool, with most studies rated as low risk.

Limitations

The methodological heterogeneity of the included studies limits the possibility of pooling effect sizes in meta-analyses.

Participant Demographics

The review included studies with a total of 1889 patients.

Statistical Information

P-Value

p<0.001

Confidence Interval

3.10% ± 95.11%

Digital Object Identifier (DOI)

10.1093/bjro/tzae041