Optimizing Radiation Dose and Image Quality in Stroke CT Protocols: Proposed Diagnostic Reference Levels for Multiphase CT Angiography and Perfusion Imaging
2024

Optimizing Radiation Dose and Image Quality in Stroke CT Protocols

Sample size: 1790 publication Evidence: high

Author Information

Author(s): Forbrig Robert, Trumm Christoph G., Reidler Paul, Kunz Wolfgang G., Dimitriadis Konstantinos, Kellert Lars, Rückel Johannes, Liebig Thomas, Stahl Robert, Abuzaid Mohamed

Primary Institution: Institute for Diagnostic and Interventional Neuroradiology, LMU University Hospital, LMU Munich

Hypothesis

What are the diagnostic reference levels for multiphase CT angiography and perfusion imaging in stroke protocols?

Conclusion

The study found that adding multiphase CT angiography scans results in only a minor increase in radiation exposure, especially with dual-source CT scanners.

Supporting Evidence

  • Local diagnostic reference levels for NCCT ranged from 37.3 to 49.1 mGy.
  • CT perfusion showed a 28.2% dose reduction with protocol adjustment.
  • Subjective image quality was rated excellent except for slightly increased noise in one scanner.

Takeaway

Doctors can use a special type of CT scan for strokes that helps them see better without giving patients too much radiation.

Methodology

A retrospective single-center study analyzing dosimetry and image quality for 1790 patients using three different CT scanners.

Limitations

The study is limited by its single-center design and the use of only one manufacturer's CT scanners.

Participant Demographics

1790 patients, 883 female, ages 18 to 99, median age 78 years.

Statistical Information

P-Value

p < 0.001

Statistical Significance

p < 0.001

Digital Object Identifier (DOI)

10.3390/diagnostics14242866