Visibility of Alveolar Bone Thickness on CBCT Images
Author Information
Author(s): Lennholm Camilla, Andreasen Hanna, Westerlund Anna, Lund Henrik
Primary Institution: Department of Oral and Maxillofacial Radiology, Institute of Odontology, Sahlgrenska Academy, Gothenburg University
Hypothesis
At which thickness does marginal bone become visible on CBCT images, and how do reconstruction techniques and viewing modes affect this assessment?
Conclusion
The best CBCT settings for analyzing alveolar bone are high-resolution images viewed in grey scale, which require greater thickness for visibility on the lingual side compared to the buccal side.
Supporting Evidence
- MPR grey scale images at high resolution showed the highest validity for bone visibility.
- Bone thickness required for visibility was 0.173 mm buccally and 0.356 mm lingually.
- 3D-rendered images had lower validity for measuring thin bony structures.
- Significant differences were found between standard and high-resolution protocols.
Takeaway
To see the bone in dental images, it needs to be thicker on the inside of the mouth than on the outside, and certain ways of taking pictures work better than others.
Methodology
Fourteen anterior teeth from six human mandibles were examined using two CBCT resolution protocols and various reconstruction techniques.
Potential Biases
Potential inter-observer variability was not assessed due to the single observer design.
Limitations
The study was conducted in vitro, which may not fully replicate clinical conditions, and was evaluated by a single observer.
Participant Demographics
Six human mandibles were used for the study.
Statistical Information
P-Value
p<0.05
Statistical Significance
p<0.05
Digital Object Identifier (DOI)
Want to read the original?
Access the complete publication on the publisher's website