The concurrent validity of a portable ultrasound probe for muscle thickness measurements
2025

Validity of a Portable Ultrasound Probe for Measuring Muscle Thickness

Sample size: 18 publication Evidence: moderate

Author Information

Author(s): Homer Kai A., Cross Matt R., Jukic Ivan

Primary Institution: Auckland University of Technology

Hypothesis

Can a portable ultrasound probe accurately measure muscle thickness compared to a stationary device?

Conclusion

The Lumify probe can be used interchangeably with the Vivid S5 for muscle thickness measurements.

Supporting Evidence

  • Measurements of muscle thickness were significantly different only at the anterior thigh.
  • Linear associations were found between the devices at each site of measurement.
  • Systematic bias was low at all sites (≤0.11 cm).
  • Proportional bias was detected only at the posterior lower leg.

Takeaway

Researchers tested a new portable ultrasound device to see if it works as well as a bigger, stationary one for measuring muscle thickness, and it does!

Methodology

Participants had their muscle thickness measured at five sites using both a portable and a stationary ultrasound device in a repeated-measures design.

Potential Biases

Proportional bias was detected at the posterior lower leg, but the absolute differences were trivial.

Limitations

The study only included young, healthy adults engaged in regular resistance training, which may limit generalizability to other populations.

Participant Demographics

18 participants (9 males and 9 females) aged 25.44 years on average.