Effective x-ray attenuation coefficient measurements from two full field digital mammography systems for data calibration applications
2008

Measuring X-ray Attenuation in Digital Mammography

publication 10 minutes Evidence: moderate

Author Information

Author(s): John J Heine, Jerry A Thomas

Primary Institution: Moffitt Cancer Center

Hypothesis

Can effective x-ray attenuation coefficients be accurately measured from full field digital mammography systems for breast density calibration?

Conclusion

The calibration quantities from FFDM systems can be used for data acquired on similar systems.

Supporting Evidence

  • Breast density is a significant breast cancer risk factor.
  • Calibration data collection can be reduced by applying a simple normalization technique.
  • Intrasystem calibration variation was on the order of four percent.

Takeaway

This study looks at how to measure breast density using digital mammography, which helps in understanding breast cancer risk.

Methodology

Logarithmic response calibration curves and effective x-ray attenuation coefficients were measured from two FFDM systems using breast tissue equivalent phantoms.

Potential Biases

Potential bias due to variations in detector response and calibration methods.

Limitations

The study is limited to two similar systems and does not assess natural serial variation in regression parameters.

Statistical Information

P-Value

0.56

Confidence Interval

0.784–0.882

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1186/1475-925X-7-13