In vivo myograph measurement of muscle contraction at optimal length
2007

Assessment of a New Myograph for Muscle Force Measurement

Sample size: 410 publication Evidence: high

Author Information

Author(s): Rahe-Meyer Niels, Winterhalter Michael, Ahmed Aminul I, Weilbach Christian, Gross Matthias, Piepenbrock Siegfried, Pawlak Matthias

Primary Institution: Hannover Medical School

Hypothesis

The new myograph can accurately measure muscle force and its reproducibility in clinical settings.

Conclusion

The new myograph is a highly reliable measuring device for assessing muscle contractions at optimal length.

Supporting Evidence

  • The myograph showed a precision of 98.74% in muscle force measurements.
  • Only two consecutive measurements are needed to assess a real, representative individual value of muscle force.
  • The mean value of muscle force was found to be 9.51 N with a 95% reference interval of 4.77–14.25 N.

Takeaway

The new myograph can measure how strong your muscles are in a very accurate way, even if you're not able to squeeze hard.

Methodology

The study involved 500 in-vivo measurements of muscle contractions on 5 test subjects and 405 healthy subjects using a newly developed myograph.

Limitations

The study was limited to healthy subjects and may not fully represent muscle force measurements in patients with neuromuscular diseases.

Participant Demographics

5 healthy voluntary test subjects (2 women, 3 men) aged 20-30 and 405 healthy subjects aged 22-41.

Statistical Information

P-Value

p<0.05

Confidence Interval

4.77–14.25 N

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1186/1475-925X-6-1

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication