Custom-designed orthopedic implants evaluated using finite element analysis
Author Information
Author(s): Harrysson Ola LA, Hosni Yasser A, Nayfeh Jamal F
Primary Institution: North Carolina State University
Hypothesis
The proposed bone-implant interface surface will provide more even stress distribution.
Conclusion
The proposed custom femoral component design reduces the risk of premature loosening and minimizes the need for additional surgical interventions.
Supporting Evidence
- The proposed design shows a more even stress distribution on the bone-implant interface surface.
- About 40% less bone must be removed with the custom design.
- The custom design can accommodate anatomical abnormalities at the distal femur.
Takeaway
This study shows that making knee implants fit better to a person's unique bone shape can help them last longer and reduce pain.
Methodology
Finite Element Analysis was used to evaluate and compare the proposed design of a custom femoral component with a conventional design based on CT scans.
Limitations
The study acknowledges the time and cost required for custom design and the potential need for surgical robots.