Comparative evaluation of the performance of the Abbott RealTime HIV-1 assay for measurement of HIV-1 plasma viral load on genetically diverse samples from Greece
2011

Comparing HIV-1 Viral Load Assays in Greece

Sample size: 310 publication 10 minutes Evidence: high

Author Information

Author(s): Katsoulidou Antigoni, Rokka Chrysoula, Issaris Catherine, Haida Catherine, Tzannis Kimon, Sypsa Vana, Detsika Maria, Paraskevis Dimitrios, Hatzakis Angelos

Primary Institution: National Retrovirus Reference Center, Department of Hygiene and Epidemiology and Medical Statistics, Athens University Medical School

Hypothesis

The study aims to compare the performance of the Abbott RealTime HIV-1 assay with other viral load assays in the context of genetically diverse HIV-1 samples.

Conclusion

The Abbott RealTime assay showed a high correlation and good agreement with the Cobas TaqMan and bDNA 3.0 assays for quantifying HIV-1 viral load across various subtypes.

Supporting Evidence

  • The Abbott RealTime assay showed a high correlation with the Cobas TaqMan assay (r = 0.960).
  • 85% of samples showed differences of less than 0.5 log10 copies/ml between Abbott RealTime and Cobas TaqMan.
  • The majority of samples (86%) differed by less than 0.5 log10 copies/ml when comparing Abbott RealTime and bDNA 3.0.

Takeaway

This study looked at different tests to measure HIV in blood and found that they mostly agree, but sometimes they give different results.

Methodology

The study compared the Abbott RealTime assay with Cobas TaqMan and bDNA 3.0 assays using clinical samples from HIV-1 positive patients.

Potential Biases

Potential mismatches in primer/probe binding sites could lead to discrepancies in viral load quantification.

Limitations

The study did not control for operator variability or test lot differences, and some samples had low viral loads which may affect results.

Participant Demographics

Samples were from HIV-1 positive patients in Greece, with various viral load levels and subtypes.

Statistical Information

P-Value

<0.001

Confidence Interval

95% limits of agreement: -0.790 to 0.379 for Abbott RealTime vs Cobas TaqMan; -0.402 to 0.721 for Abbott RealTime vs bDNA 3.0

Statistical Significance

p<0.001

Digital Object Identifier (DOI)

10.1186/1743-422X-8-10