Turbulent flow as a cause for underestimating coronary flow reserve measured by Doppler guide wire
2006

Turbulent Flow and Coronary Flow Reserve

Sample size: 75 publication Evidence: moderate

Author Information

Author(s): Markus Ferrari, Gerald S. Werner, Philipp Bahrmann, Barbara M. Richartz, Hans R. Figulla

Primary Institution: Friedrich-Schiller-University, Jena, Germany

Hypothesis

Turbulent flow may occur in human coronary arteries influencing Doppler flow measurements.

Conclusion

Doppler guide wires are valid for measuring coronary flow velocity below a critical Reynolds number of 500, but may underestimate flow by up to 22.5% in turbulent conditions.

Supporting Evidence

  • Doppler wires showed a significant correlation to perfusion volumes under laminar flow conditions.
  • Above a critical Reynolds number of 500, flow measurements underestimated actual perfusion rates.
  • In 20% of patients, maximum flow velocities exceeded the critical Reynolds number.

Takeaway

Doctors use special wires to measure blood flow in the heart, but if the blood is swirling too much, they might think there's less flow than there really is.

Methodology

In vitro and in vivo studies using Doppler-tipped guide wires to measure coronary flow velocity in glass pipes and patients.

Potential Biases

Potential underestimation of flow due to turbulent conditions not accounted for in all patients.

Limitations

The study's in vitro conditions do not fully replicate the complexities of human coronary arteries.

Participant Demographics

75 patients, average age 59.5 years, 77.3% men.

Statistical Information

P-Value

p<0.05

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1186/1476-7120-4-14

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication