Comparison of lens- and fiber-coupled CCD detectors for X-ray computed tomography
2011

Comparison of X-ray Detectors for Computed Tomography

publication Evidence: moderate

Author Information

Author(s): K. Uesugi, M. Hoshino, N. Yagi

Primary Institution: Japan Synchrotron Radiation Research Institute

Hypothesis

How do lens-coupled and fiber-coupled CCD detectors compare in performance for X-ray computed tomography?

Conclusion

The fiber-coupled detector is about four times more efficient than the lens-coupled detector in light transmission, making it advantageous for fast data collection in X-ray imaging.

Supporting Evidence

  • The fiber-coupled detector has a transmittance of light to the CCD that is about four times higher than the lens-coupled detector.
  • The lens-coupled detector shows a global shading of up to 40% in response uniformity.
  • The fiber-coupled detector is more suitable for high-speed tomography of soft materials.

Takeaway

This study looked at two types of X-ray detectors to see which one works better. The fiber-coupled one is better at letting light through, which helps take pictures faster.

Methodology

The performance of lens- and fiber-coupled CCD detectors was compared in high-resolution computed tomography experiments at the SPring-8 synchrotron radiation facility.

Limitations

The fiber-coupled detector has issues with browning of the fiber and image distortion, while the lens-coupled detector suffers from shading effects.

Digital Object Identifier (DOI)

10.1107/S0909049510044523