Comparison of X-ray Detectors for Computed Tomography
Author Information
Author(s): K. Uesugi, M. Hoshino, N. Yagi
Primary Institution: Japan Synchrotron Radiation Research Institute
Hypothesis
How do lens-coupled and fiber-coupled CCD detectors compare in performance for X-ray computed tomography?
Conclusion
The fiber-coupled detector is about four times more efficient than the lens-coupled detector in light transmission, making it advantageous for fast data collection in X-ray imaging.
Supporting Evidence
- The fiber-coupled detector has a transmittance of light to the CCD that is about four times higher than the lens-coupled detector.
- The lens-coupled detector shows a global shading of up to 40% in response uniformity.
- The fiber-coupled detector is more suitable for high-speed tomography of soft materials.
Takeaway
This study looked at two types of X-ray detectors to see which one works better. The fiber-coupled one is better at letting light through, which helps take pictures faster.
Methodology
The performance of lens- and fiber-coupled CCD detectors was compared in high-resolution computed tomography experiments at the SPring-8 synchrotron radiation facility.
Limitations
The fiber-coupled detector has issues with browning of the fiber and image distortion, while the lens-coupled detector suffers from shading effects.