Evaluating a New Software for Measuring Carotid Intima-Media Thickness
Author Information
Author(s): Potter Kathleen, Green Daniel J, Reed Christopher J, Woodman Richard J, Watts Gerald F, McQuillan Brendan M, Burke Valerie, Hankey Graeme J, Arnolda Leonard F
Primary Institution: School of Medicine and Pharmacology, University of Western Australia, Perth, Australia
Hypothesis
Increasing the number of measurements used to estimate CIMT will reduce random measurement error and improve reproducibility.
Conclusion
The new DICOM-based software significantly reduced analysis time, but increasing the number of measurements did not improve reproducibility.
Supporting Evidence
- The new software halved CIMT measurement time compared to the alternative system.
- A ten-fold increase in measurements did not improve reproducibility.
- Observer biases were more pronounced with the sixty-frame protocol.
Takeaway
The new software makes measuring a part of the neck easier and faster, but taking more pictures doesn't make the results more reliable.
Methodology
Two sonographers recorded carotid scans in thirty volunteers, and CIMT was measured using both new and old software.
Potential Biases
Observer biases were present, particularly with the sixty-frame measurement protocol, which increased the risk of bias.
Limitations
One observer did not perform replicate measures with the multiple-frame protocol, reducing the ability to detect improvements in intra-observer reproducibility.
Participant Demographics
{"normal":8,"hyperlipidaemic":7,"stroke":15,"age_mean":{"normal":38,"hyperlipidaemic":57,"stroke":64},"male_count":{"normal":2,"hyperlipidaemic":6,"stroke":11}}
Statistical Information
P-Value
p < 0.0001
Confidence Interval
95%CI 0.01–0.03 mm
Statistical Significance
p < 0.0001