A rare complication of intracardiac double knotting of temporary pacemaker lead during bedside insertion: a case report
2024

Rare Complication of Temporary Pacemaker Lead Knotting

Sample size: 1 publication Evidence: low

Author Information

Author(s): Aditi Dattagupta, Shweta Agrawal, Srilakshmi Adhyapak, Harshith Kramadhari, Abhilash Konda

Primary Institution: St. John’s National Academy of Health Sciences

Conclusion

Intracardiac knotting of temporary pacing leads is rare but can be successfully retrieved without surgery by using a snare technique.

Supporting Evidence

  • Intracardiac knotting of temporary pacing leads is a rare complication.
  • The retrieval of the knotted lead was successfully performed without surgical intervention.
  • Using echocardiography guidance can help in correct placement of the temporary lead.

Takeaway

Sometimes, when doctors put in a temporary pacemaker, the wire can get tangled in knots. This case shows how to safely get it out without needing surgery.

Methodology

The case involved a patient with a knotted temporary pacemaker lead, which was retrieved using a snare technique through a femoral sheath.

Potential Biases

Potential bias due to the lack of multiple cases or comparative studies.

Limitations

The case report is based on a single patient, limiting generalizability.

Participant Demographics

A 47-year-old male patient.