New Method for Focused Ultrasound Treatment
Author Information
Author(s): Kong Dechen, Liu Gaojie, Cheng Bochao, Qi Xu, Zhu Jiayu, He Qiang, Xing Haoyang, Gong Qiyong
Primary Institution: College Of Physics, Sichuan University, Chengdu, China
Hypothesis
Using UTE images as the prDeep initial input can drastically reduce the number of iterations required for the calibration of TUS therapy.
Conclusion
The hybrid method significantly reduces the number of iterations needed for transcranial ultrasound therapy while maintaining accuracy and robustness.
Supporting Evidence
- The prDeep algorithm showed an average correlation of approximately 0.90 and an average MSE lower than 0.018.
- The method restored 75% of the gold standard intensity in 25 iteration steps.
- The hybrid technique demonstrated excellent adaptability and anti-noise capacity.
Takeaway
This study shows a new way to use sound waves to treat brain problems, making it faster and better at focusing on the right spot.
Methodology
The study used a hybrid method combining ultrashort echo time MRI sequences with phase retrieval techniques to improve ultrasound focusing.
Potential Biases
The study did not use hydrophones for ex vivo human skull TM measurement due to hardware limitations.
Limitations
The sample size was relatively small, and the DnCNN was not specially trained.
Participant Demographics
Three physically active subjects with no history of nervous system disease participated.
Statistical Information
P-Value
p<0.05
Statistical Significance
p<0.05
Digital Object Identifier (DOI)