Method for Measuring Nebivolol in Tablets
Author Information
Author(s): Shah D. A., Bhatt K. K., Mehta R. S., Baldania S. L., Gandhi T. R.
Primary Institution: Anand Pharmacy College
Hypothesis
The study aims to develop a new, selective and sensitive liquid chromatographic method for the estimation of nebivolol in tablets for routine quality control analysis and stability studies.
Conclusion
The proposed method for estimating nebivolol hydrochloride in tablet formulations is simple, sensitive, accurate, and precise.
Supporting Evidence
- The method showed a linearity range for nebivolol from 0.2 to 10 μg/ml.
- The recovery of nebivolol was found to be between 98.68% and 100.86%.
- The detection limit was established at 0.06 μg/ml and the quantification limit at 0.2 μg/ml.
- The method was validated for specificity, accuracy, and precision.
- Degradation studies indicated that nebivolol is susceptible to hydrolysis and oxidation.
Takeaway
Researchers created a new way to measure a heart medication in tablets to make sure it's safe and effective.
Methodology
A reversed phase liquid chromatographic method was developed using a Phenomenex Gemini C-18 column and a specific mobile phase for the analysis of nebivolol.
Limitations
The study does not mention specific limitations.