Assessing Gingival Thickness with Different Methods
Author Information
Author(s): Gkogkos Andreas, Kloukos Dimitrios, Koukos George, Liapis George, Sculean Anton, Katsaros Christos
Primary Institution: Department of Periodontology, 251 Hellenic Air Force & VA General Hospital, Athens, Greece; School of Dental Medicine, University of Bern, Switzerland
Hypothesis
The study aims to assess the agreement between various clinical and radiographic methods for evaluating gingival thickness.
Conclusion
CBCT measurements proved to be highly repeatable and comparable to the USD measurements, while both CBCT and USD measurements were systematically higher than either PB or AN.
Supporting Evidence
- Trans-gingival measurements with the standard probe and acupuncture needle were almost identical.
- Ultrasound and CBCT yielded statistically significantly higher values than the acupuncture needle.
- There was no evidence of systematic differences between repeated CBCT measurements.
Takeaway
The study looked at how thick gums are using different tools, finding that some tools give similar results while others show thicker gums.
Methodology
This ex-vivo study evaluated gingival thickness on 20 porcine cadavers using four methods: trans-gingival probing with a standard probe, an acupuncture needle, ultrasound, and Cone Beam Computed Tomography.
Potential Biases
Potential bias due to the subjective nature of visual assessments and the invasive nature of some measurement methods.
Limitations
The results may not be applicable to all tooth areas in human subjects due to anatomical differences and the subjective nature of visual assessments.
Participant Demographics
20 porcine cadavers, 6 months old.
Statistical Information
P-Value
0.06 for left incisor, 0.55 for right incisor
Confidence Interval
95% CI = -0.10, 0.01 for left incisor; 95% CI = -0.07, 0.04 for right incisor
Statistical Significance
p<0.05
Digital Object Identifier (DOI)