Reproducibility of Daily Isocenter Verification in Prostate Cancer Treatment
Author Information
Author(s): Ullman Karen L, Ning Holly, Susil Robert C, Ayele Asna, Jocelyn Lucresse, Havelos Jan, Guion Peter, Xie Huchen, Li Guang, Arora Barbara C, Cannon Angela, Miller Robert W, Norman Coleman C, Camphausen Kevin, Ménard Cynthia
Primary Institution: Radiation Oncology Branch, Center for Cancer Research, National Cancer Institute, NIH, DHHS
Hypothesis
We sought to determine the intra- and inter-radiation therapist reproducibility of a previously established matching technique for daily verification and correction of isocenter position relative to intraprostatic fiducial markers.
Conclusion
We have found high reproducibility with a previously established method for daily verification and correction of isocenter position relative to prostatic fiducial markers using electronic portal imaging.
Supporting Evidence
- The mean intra-observer variability ranged from 0.4 to 0.7 mm.
- Inter-observer results showed a mean variability of 0.9 mm.
- Only 0.5% of treatments required a table shift due to intra or inter-observer error at a 5 mm threshold.
- The error rate increased to 2.4% if the threshold was reduced to 3 mm.
- A statistically significant difference was found in observer variability between lateral and AP portal image manual matches.
Takeaway
This study shows that radiation therapists can consistently check and adjust the position of prostate cancer treatment equipment, which helps ensure patients receive the correct dose of radiation.
Methodology
Four therapists repeated and recorded the alignment of fiducial markers on 20 previously acquired portal image datasets from two patients.
Potential Biases
Potential bias due to the short interval between measurements and the lack of assessment of rotational errors.
Limitations
The study did not assess the accuracy of the actual table shift performed by therapists and had a short interval between measurements.
Participant Demographics
Patients with localized prostate cancer undergoing external beam radiotherapy.
Statistical Information
P-Value
p<0.01
Confidence Interval
0.75–0.84
Statistical Significance
p<0.01