Are the calf and the thigh reliable alternatives to the arm for cuff non-invasive measurements of blood pressure?
2011

Measuring Blood Pressure: Arm vs. Calf and Thigh

Sample size: 150 publication Evidence: moderate

Author Information

Author(s): Lakhal K, Macq C, Ehrmann S, Boulain T, Capdevila X

Primary Institution: CHU, Montpellier, France

Hypothesis

Are the calf and the thigh reliable alternatives to the arm for cuff non-invasive measurements of blood pressure?

Conclusion

For better reliability of MAP measurements, the cuff should be placed on the arm rather than the thigh or the calf.

Supporting Evidence

  • NIBP performed better on the arm compared to the calf and thigh.
  • Accuracy of NIBP was similar in cases of mild circulatory failure.
  • MAP changes were better reflected when the cuff was placed on the arm.

Takeaway

This study found that using the arm to measure blood pressure is more accurate than using the calf or thigh.

Methodology

Adult ICU patients with arterial catheters were included, and NIBP measurements were compared across different anatomical sites.

Limitations

Some patients were excluded due to measurement failures, and the study focused only on ICU patients.

Participant Demographics

150 patients, average age 41 years, BMI 26, majority on mechanical ventilation.