The establishment of reference intervals for the ClotPro thromboelastometry device in healthy dogs
2024

Reference Intervals for a Thromboelastometry Device in Healthy Dogs

Sample size: 48 publication Evidence: moderate

Author Information

Author(s): Day Samantha K., Nash Katherine J., Midwinter Mark J., Purcell Sarah L., Goodwin Wendy A.

Primary Institution: The University of Queensland

Hypothesis

Increasing storage time would increase the hypercoagulability of the tracing.

Conclusion

The development of reference intervals for the thromboelastometry device allows for the clinical use of this technology, but prolonged storage times may lead to erroneous results.

Supporting Evidence

  • Reference intervals were calculated using robust methods for parametric data.
  • Increasing storage time resulted in more hypocoagulable tracings.
  • A correlation was found between the IN-test and aPTT.
  • Statistically significant correlations were classified as weak or moderate.

Takeaway

This study helps veterinarians understand how to use a new blood test for dogs, showing that the test results can change if the blood is stored too long.

Methodology

Forty-eight healthy dogs were enrolled, and various thromboelastometry tests were performed on blood samples collected at different storage times.

Potential Biases

Convenience sampling may have introduced bias despite adherence to inclusion and exclusion criteria.

Limitations

The study had a small sample size and some tests were excluded due to errors, which may limit the generalizability of the results.

Participant Demographics

47 dogs included, 25 females (20 neutered, 5 entire) and 22 males (17 neutered, 5 entire), median age 4.75 years, median weight 21.0 kg.

Statistical Information

P-Value

p<0.0001

Statistical Significance

p<0.0001

Digital Object Identifier (DOI)

10.1111/vec.13426

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication