Focus on POCUS: Identification of Early Successful Intubation by Point-of-Care Ultrasound Versus End-Tidal Carbon Dioxide: A Prospective Comparative Study
2024

Using Ultrasound to Confirm Successful Intubation Faster than Capnography

Sample size: 104 publication Evidence: high

Author Information

Author(s): Neethirajan Soma Ganesh Raja, Baskar Ganeshamoorthy, Parameswari Aruna

Primary Institution: Sri Ramachandra Institute of Higher Education and Research

Hypothesis

Can ultrasound provide a quicker confirmation of successful endotracheal intubation compared to end-tidal carbon dioxide monitoring?

Conclusion

Ultrasound is a faster and more reliable tool for confirming successful endotracheal intubation than end-tidal carbon dioxide monitoring.

Supporting Evidence

  • Ultrasound confirmed successful intubation in an average of 21.63 seconds.
  • Capnography required an average of 40.62 seconds for confirmation.
  • The mean difference in recognition time between ultrasound and capnography was 18.98 seconds.
  • Ultrasound is particularly useful in high-risk emergency situations.
  • Ultrasound can detect misdirected intubation in real-time.

Takeaway

Doctors can use ultrasound to check if a patient is properly intubated much faster than using the traditional method of checking carbon dioxide levels.

Methodology

A prospective observational study comparing the time taken for ultrasound and capnography to confirm successful intubation in 104 patients.

Limitations

The study did not statistically address the early identification of esophageal intubation.

Participant Demographics

Patients aged 18-75 years, with a mix of male (56.73%) and female (43.27%) participants.

Statistical Information

P-Value

<0.001

Confidence Interval

18.15-19.81

Statistical Significance

p<0.001

Digital Object Identifier (DOI)

10.4274/TJAR.2024.241720