Investigation of the hydrodynamic properties of a new MRI-resistant programmable hydrocephalus shunt
2008

Testing a New Hydrocephalus Shunt

Sample size: 3 publication Evidence: moderate

Author Information

Author(s): Allin David M, Czosnyka Marek, Richards Hugh K, Pickard John D, Czosnyka Zofia H

Primary Institution: Shunt Evaluation Laboratory & Academic Neurosurgical Unit, Addenbrooke's Hospital, Cambridge, UK

Hypothesis

The study aims to measure the properties of the Polaris valve to provide independent data about its performance.

Conclusion

The Polaris Valve is a reliable, adjustable valve that cannot be accidentally re-adjusted by an external magnetic field.

Supporting Evidence

  • The Polaris valve showed good mechanical durability over a 3-month testing period.
  • Hydrodynamic resistance varied with the performance level, ranging from 1.6 to 11.2 mmHg/ml/min.
  • The valve's programming was unaffected by external magnetic fields up to 3 Tesla.

Takeaway

The Polaris valve is a new device that helps drain fluid from the brain and works well without being affected by strong magnets.

Methodology

The performance and hydrodynamic properties of the Polaris valve were evaluated in a laboratory setting using various pressure and flow tests.

Potential Biases

The study was commissioned by the manufacturer, which may introduce bias.

Limitations

The study was conducted in vitro, and clinical outcomes may differ.

Statistical Information

P-Value

p<0.05

Statistical Significance

p<0.05

Digital Object Identifier (DOI)

10.1186/1743-8454-5-8