Comparison of cardiac index: LiDCOrapid and PiCCOplus in the ICU
2011

Comparison of Cardiac Index Measurement Methods in ICU Patients

Sample size: 30 publication Evidence: moderate

Author Information

Author(s): Brass P, Mills E, Latza J, Peters J, Berendes E

Primary Institution: Helios Klinikum, Krefeld, Germany

Hypothesis

This study aims to compare two arterial pressure wave-form monitors for cardiac index measurement.

Conclusion

LiDCOrapid CI values are in agreement with PiCCO CI values, showing excellent concordance and minimal bias.

Supporting Evidence

  • The study included 30 patients with a wide age range.
  • Regression analysis showed a high degree of correlation between the two measurement methods.
  • Concordance analysis indicated 97.8% overall agreement for changes in cardiac index.

Takeaway

The study looked at two ways to measure heart function in sick patients and found they work well together.

Methodology

20 patients who had major surgery and 10 with circulatory instability were monitored using two different devices.

Participant Demographics

Patients aged 31 to 90, with varying ASA classifications.

Digital Object Identifier (DOI)

10.1186/cc9482

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication