Light-guided dynamic phantom to mimic microvasculature for biomedical applications: an exploration for pulse oximeter
2024

Light-guided dynamic phantom for pulse oximetry

publication Evidence: high

Author Information

Author(s): Hui Ma, Dario Angelone, Claudia Nunzia Guadagno, Stefan Andersson-Engels, Sanathana Konugolu Venkata Sekar

Primary Institution: Tyndall National Institute, Cork, Ireland

Hypothesis

We aimed to develop a solid dynamic phantom that can rapidly change optical properties to simulate biological signals.

Conclusion

The dynamic phantom can simulate various biological signals and has potential applications in calibrating and validating pulse oximeters and imaging systems.

Supporting Evidence

  • The dynamic phantom can generate photoplethysmography signals at different heart rates.
  • It successfully simulates oxygen saturation levels at various percentages.
  • The phantom's design allows for real-time control of optical properties.
  • It can be used to calibrate and validate pulse oximeters and imaging systems.

Takeaway

This study created a special tool that can mimic how blood behaves in the body, helping doctors use devices that measure oxygen levels more accurately.

Methodology

The phantom's absorption properties were controlled in real time using an LCD, and it was characterized at four wavelengths: 940, 660, 530, and 455 nm.

Limitations

The phantom cannot simulate vascular compliance during the cardiac cycle due to its fixed geometric structure.

Digital Object Identifier (DOI)

10.1117/1.JBO.29.S3.S33312

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication