Level of Agreement Between Plaque Detection with Clinical Assessment and Assessment on Intraoral Scanner
2024

Agreement Between Intraoral Scanner and Clinical Method for Plaque Detection

Sample size: 14 publication Evidence: moderate

Author Information

Author(s): Gkavela Grigoria, Nørrisgaard Pia Elisabeth, Rahiotis Christos

Primary Institution: National and Kapodistrian University of Athens

Hypothesis

The study aims to evaluate the agreement between plaque detection using an intraoral scanner and a conventional clinical method.

Conclusion

The level of agreement between conventional clinical registration and registration from 3D models was acceptable overall.

Supporting Evidence

  • All methods showed moderate to strong correlations.
  • Cronbach’s α ranged from 0.551 to 0.766.
  • The ICC value was calculated to be 0.82, indicating good reliability.
  • Spearman’s rho values indicated moderate to strong agreement.

Takeaway

This study looked at how well a new scanning method for detecting plaque in teeth works compared to the traditional way, and it found that they agree pretty well.

Methodology

Participants were clinically examined for plaque using the modified Quigley–Hein plaque index and scanned with an intraoral scanner before and after using a disclosing agent.

Potential Biases

Subjectivity in plaque scoring may introduce bias.

Limitations

The small sample size limits the generalizability of the findings.

Participant Demographics

14 participants (8 female, 6 male) with a mean age of 37.5 years.

Statistical Information

P-Value

<0.001

Statistical Significance

p<0.001

Digital Object Identifier (DOI)

10.3390/dj12120395