Evaluation of the Olympus AU-510 analyser
1991

Evaluation of the Olympus AU-510 Analyser

Sample size: 12 publication Evidence: moderate

Author Information

Author(s): Carmen Farr, Jestls Velasco, Francisco Ram6n

Primary Institution: Servei de Bioquimica, Hospital SantJoande Deu

Conclusion

The Olympus AU-510 analyser demonstrated acceptable imprecision and accuracy for various serum analytes under routine conditions.

Supporting Evidence

  • The within-run imprecision ranged from 0.67% for phosphate to 2.89% for iron.
  • There was no carry-over in the serum sample pipetter.
  • Carry-over studies showed some cross-contamination in the iron assay.

Takeaway

The Olympus AU-510 is a machine that helps test blood samples, and it works pretty well for different tests.

Methodology

The evaluation included a photometric study and an assessment of within-run and between-run imprecision for 12 serum analytes.

Limitations

The results cannot be compared with other instruments due to varying methodologies.