Evaluation of the Olympus AU-510 analyser
1991
Evaluation of the Olympus AU-510 Analyser
Sample size: 12
publication
Evidence: moderate
Author Information
Author(s): Carmen Farr, Jestls Velasco, Francisco Ram6n
Primary Institution: Servei de Bioquimica, Hospital SantJoande Deu
Conclusion
The Olympus AU-510 analyser demonstrated acceptable imprecision and accuracy for various serum analytes under routine conditions.
Supporting Evidence
- The within-run imprecision ranged from 0.67% for phosphate to 2.89% for iron.
- There was no carry-over in the serum sample pipetter.
- Carry-over studies showed some cross-contamination in the iron assay.
Takeaway
The Olympus AU-510 is a machine that helps test blood samples, and it works pretty well for different tests.
Methodology
The evaluation included a photometric study and an assessment of within-run and between-run imprecision for 12 serum analytes.
Limitations
The results cannot be compared with other instruments due to varying methodologies.