Pulse Contour Analysis After Cardiac Surgery
Author Information
Author(s): Sander Michael, Heymann Christian, Foer Achim, von Dossow Vera, Grosse Joachim, Dushe Simon, Konertz Wolfgang F, Spies Claudia D
Primary Institution: University Hospital Charité, Campus Charité Mitte, University Medicine, Berlin, Germany
Hypothesis
Is the continuous arterial pulse contour analysis accurate after cardiopulmonary bypass in cardiac surgery patients?
Conclusion
Pulse contour analysis is not reliable after cardiopulmonary bypass and should be recalibrated to avoid inaccurate results.
Supporting Evidence
- Prior to CPB, there was an excellent correlation between COPACtherm and COPiCCOtherm measurements.
- After CPB, pulse contour analysis showed marked differences compared to thermodilution methods.
- Recalibration of pulse contour analysis is essential after cardiopulmonary bypass.
Takeaway
Doctors used a special method to measure heart output during heart surgery, but it didn't work well after the surgery, so they need to fix it quickly.
Methodology
This was a prospective study analyzing data from 45 patients undergoing coronary artery bypass graft surgery, comparing two methods of cardiac output measurement.
Potential Biases
The study design does not allow for a definitive causal relationship between cardiopulmonary bypass and measurement inaccuracies.
Limitations
The study did not use a method independent of thermal signals to validate the measurements.
Participant Demographics
Patients aged 18-75 undergoing elective coronary artery bypass graft surgery.
Statistical Information
P-Value
<0.001
Statistical Significance
p<0.001
Digital Object Identifier (DOI)
Want to read the original?
Access the complete publication on the publisher's website