Calculating Optic Nerve Margins for Stereotactic Radiosurgery
Author Information
Author(s): Sabharwal Sagar MSc, Heyes Geoff PhD, Tudor George S J PhD, Flintham Robert MSc, Chavda Swarupsinh MD, Sanghera Paul MBChB
Primary Institution: University Hospitals Birmingham NHS Foundation Trust
Hypothesis
The study aims to calculate optic nerve planning organ at risk volume (PRV) margins for CyberKnife SRS treatment planning based on optic nerve motion measured using MRI.
Conclusion
Optic nerve PRV margins may be needed in CyberKnife planning to reduce risk to the optic nerves.
Supporting Evidence
- Optic nerve motion can significantly affect radiation dose during treatment.
- Measured optic nerve motion was used to calculate PRV margins for CyberKnife SRS.
- Two types of PRV margins were calculated based on different patient gaze scenarios.
Takeaway
Doctors need to be careful about how the optic nerves move during treatment, so they calculated special safety margins to protect them.
Methodology
MRI scans were taken of 10 healthy volunteers to measure optic nerve motion, which was then used to calculate PRV margins.
Potential Biases
Potential bias due to the small sample size and the assumption regarding the distribution of optic nerve motion.
Limitations
The study is limited by the small sample size of healthy volunteers and the assumption that optic nerve motion is normally distributed.
Participant Demographics
Healthy volunteers, specific demographics not detailed.