Recovery of Perfluorocarbon Liquid During Partial Liquid Ventilation
Author Information
Author(s): Dunster Kimble R, Davies Mark W, Fraser John F
Primary Institution: Royal Women's Hospital, Brisbane, Queensland, Australia
Hypothesis
How much perfluorocarbon liquid can be recovered from gases during partial liquid ventilation, and does the recovery differ with flow rate?
Conclusion
Using two condensers, 47% to 91% of perfluorocarbon liquid can be recovered from gases during partial liquid ventilation.
Supporting Evidence
- 47% to 91% of the infused perfluorocarbon liquid was recovered at different flow rates.
- Significantly higher recovery was achieved at lower flow rates.
- The study demonstrated recovery of perfluorocarbon from expired gases that are not saturated with PFC.
Takeaway
The study shows that we can get back a lot of special liquid used in breathing treatments, which helps save money and protect the environment.
Methodology
The study used a modified ventilator circuit with two condensers to measure the recovery of perfluorocarbon liquid from expired gases at various flow rates.
Limitations
The study was limited to specific flow rates and may not represent all clinical scenarios.