An advanced expiratory circuit for the recovery of perfluorocarbon liquid from non-saturated perfluorocarbon vapour during partial liquid ventilation: an experimental model
2006

Recovery of Perfluorocarbon Liquid During Partial Liquid Ventilation

publication Evidence: moderate

Author Information

Author(s): Dunster Kimble R, Davies Mark W, Fraser John F

Primary Institution: Royal Women's Hospital, Brisbane, Queensland, Australia

Hypothesis

How much perfluorocarbon liquid can be recovered from gases during partial liquid ventilation, and does the recovery differ with flow rate?

Conclusion

Using two condensers, 47% to 91% of perfluorocarbon liquid can be recovered from gases during partial liquid ventilation.

Supporting Evidence

  • 47% to 91% of the infused perfluorocarbon liquid was recovered at different flow rates.
  • Significantly higher recovery was achieved at lower flow rates.
  • The study demonstrated recovery of perfluorocarbon from expired gases that are not saturated with PFC.

Takeaway

The study shows that we can get back a lot of special liquid used in breathing treatments, which helps save money and protect the environment.

Methodology

The study used a modified ventilator circuit with two condensers to measure the recovery of perfluorocarbon liquid from expired gases at various flow rates.

Limitations

The study was limited to specific flow rates and may not represent all clinical scenarios.

Statistical Information

P-Value

p < 0.0001

Statistical Significance

p < 0.0001

Digital Object Identifier (DOI)

10.1186/1475-925X-5-7

Want to read the original?

Access the complete publication on the publisher's website

View Original Publication